Refractometer Specifications J457 (All models)

Measurement scales:Refractive Index (nD), Brix (% Sucrose),
and up to 100 custom programmed scales
Measurement range:Refractive Index 1.32 – 1.45
Brix 0 – 100
Accuracy:Refractive Index ±0.00002
Brix 0.015
Reproducibility:Refractive Index ±0.000005
Brix 0.015
Resolution:Refractive Index 0.000001, Brix 0.01
Temperature control range:All samples are brought to 20.00°C
before measurement
Temperature control reproducibility:±0.01°C
Ambient temperature limit:5°C to 40°C
Temperature correction range:4°C to 95°C (for sucrose solutions)
Sample temperature limit:-20°C to 250°C
Optical wavelength:589.3nm (NaD line)
Response time:User configurable, generally less than
30 seconds
Calibration:Using water or NIST traceable fl uids.
Factory default calibration can always
be reset.
Prism:Artificial sapphire
Acid resistance:Hastelloy™ measurement surface (optional)
Data storage/ internal memory:32 GB Non-removable Compact Flash
Display:Adjustable 10.4 inch diagonal, 800-600
pixels, color, Flat Panel Monitor with
Resistant Touch Screen Interface, 400 nits brightness, gasketted for spill protection
User interface:Touchscreen
Communication interface:3 USB, RS232 and Cat5 Network
(Ethernet)
Operating dimensions/weight:L: 17 1/4” W: 12” H: 13” / 23 lbs.
L: 43.5cm W: 30.5cm H: 33cm / 10.4 kg
Power requirements:100 – 240 volts, 50 Hz – 60